OX-1 OxSim®

Test the complete SpO2 system, including the sensor, extension cable with the OxSim® Optical SpO2 Pulse Oximeter Simulator. Slips into your shirt pocket and operates easily with just two buttons on a single AA battery. The auto-detection of the Sensor type eliminates the need to select the manufacturer oximeter under test. The optional serial communications port enables users to expand simulation settings for Saturation, Pulse Rate, and Perfusion Index, ideal for applications where a wide range of simulations is needed.

*Custom Saturation Settings are available for Alaris, Zoll & BCI testing protocols.

OxSim® Optical SpO2 Pulse Oximeter Tester Features:

  • SpO2 Simulations:
    • Mode 1:
      • Saturation: 85%
      • Heart Rate: 80 bpm
      • Perfusion Index: approx. 2.0
    • Mode 2:
      • Saturation: 95%
      • Heart Rate: 40 bpm
      • Perfusion Index: approx. 2.0
    • Mode 3:
      • Saturation: 98%
      • Heart Rate: 80 bpm
      • Perfusion Index: approx. 2.0
    • Mode 4:
      • Saturation: 98%
      • Heart Rate: 140 bpm
      • Perfusion Index: approx. 2.0
    • Mode 5:
      • Saturation: 99%
      • Heart Rate: 80 bpm
      • Perfusion Index: approx. 0.2
  • Size:
    • 2-1/2″ x 2-1/2″ x 1-1/4″ (6.4 x 6.4 x 3.2 cm), excluding finger section
  • Weight:
    • Only 5 oz. (0.14 kg, excluding battery)
  • Simple operation:
    • Just two buttons
  • Convenient:
    • No adapter cables to purchase or carry
  • Economical:
    • ‘Finger’ is fixed and permanent. No parts to carry or replace annually
  • Rugged:
    • Polycarbonate housing. Drop tested 65 time from a height of 3 feet (1m) to a hard surface with no failures!
  • Power:
    • 6VDC Adapter or AA battery. Compatible with the Battery Boost — run OxSim & SimCube NIBP Simulator simultaneously.
  • Battery Life:
    • Minimum 8 hours. 60 days in sleep mode (unit sleeps after 10 min. with no probe connection)
  • Warranty:
    • Four-Year Limited Warranty

Package Contents:

  • (1) OxSim® Optical SpO2 Pulse Oximeter Tester
  • (1) Power Supply (6VDC)
  • (1) AA battery
  • (1) Certificate of Calibration
  • (1) Carrying Case

Optional Features:

  • Serial Communications port provides user control of OxSim for expanded simulation settings.
  • Saturation Range
    • 10-100%
  • Pulse Rate Range
    • 10-250 BPM
  • Perfusion Index Range
    • 0.1 – 2 .0 PI
×